January 2015
10
Quick Start Guide
Step 3: Verify the tag
The 848T has a removable commissioning tag that
contains both the Device ID (the unique code that
identifies a particular device in the absence of a device
tag) and a space to record the device tag (the
operational identification for the device as defined by
the Piping and Instrumentation Diagram [P&ID]).
When commissioning more than one device on a
Foundation fieldbus segment, it can be difficult to
identify which device is at a particular location. The
removable tag aids in this process by linking the Device
ID to physical location. The installer should note the
physical location of the transmitter on both the upper
and lower location of the commissioning tag. The
bottom portion should be torn off for each device on the
segment and used for commissioning the segment in
the control system.
Step 4: Verify the transmitter configuration
Each FOUNDATION fieldbus host or configuration tool has a different way of
displaying and performing configurations. Some use Device Descriptions (DD) or
DD wizards for configuration and to display data consistently across platforms.
There is no requirement that a host or configuration tool support these features.
The following is the minimum configuration requirement for a temperature
measurement. This guide is designed for systems not using DD wizards. For a
complete list of parameters and configuration information refer to the
Rosemount 848T Reference Manual (document number 00809-0100-4697).
Sensor transducer block
This block contains temperature measurement data for all eight sensor inputs. It
also includes information about sensor type, engineering units, damping, and
diagnostics. At a minimum, verify the parameters in Ta bl e 1 for each sensor in the
transducer block.
Table 1. Sensor Transducer Block Parameters
Parameter Comments
Typical Configuration N/A
Configure Input N/A
SENSOR_1_CONFIG.SENSOR example: “PT100_A_385: 3-wire
Przeglądanie stron 9